Tag Archives: Charging Station

IR Modulation Processing Algorithm Development – Part X

Posted 24 June 2017

Well, I may have spoken too soon about the perfection of my implementation of John’s ‘N-path’ band-pass filter (BPF) intended to make Wall-E2 impervious to IR interference.  After my last post on this subject, I re-ran some of the ‘Final Value’ plots for different received IR modulation amplitudes and the results were, to put it bluntly, crap 🙁 . Shown below is my original plot from yesterday, followed by the same plot for different input amplitudes

Computed final values vs complete input data cycles for sensor channel 1 (This is the original from yesterday)

 

So, clearly something is ‘fishy in Denmark’ here, when the ‘no-signal’ case with only high-frequency noise causes the output to increase without limit, and the ‘input grounded’ case is decidedly non-zero (although the values are much lower than in the ‘signal present’ cases).

Time to go back through the entire algorithm (again!) looking for the problem(s).  

25 June 2017

My original implementation of the algorithm was set up to handle four sensor input channels, so each step of the process required an iteration step to go through all four, something like the code snippet below:

In order to simplify the debug problem, I decided to eliminate all these iteration steps and just focus on one channel.  To do this I ‘branched’ my project into a ‘SingleChannel’ branch using GIT and TortoiseGit’s wonderful GIT user interface (thanks TortoiseGit!).  This allows me to muck about in a new sandbox without completely erasing my previous work – yay!

Anyway, I eliminated all the 4-sensor iteration steps, and went back through each step to make sure each was operating properly.  When I was ‘finished’ (I never really finish with any program – I just tolerate whatever level of bugs or imperfections it has for some time).  After this, I ran some tests for proper operation using just one channel.  For these tests, the Teensy ADC channel being used was a) grounded, b) connected to 3.3VDC, c) unterminated.  For each condition I captured the ‘Final Value’ output from the algorithm and plotted it in Excel, as shown below.

Single channel testing with grounded, unterminated, and +3.3VDC input

As can be seen from the above plot, things seem to be working now, at least for a single channel.  The ‘grounded’ and ‘3.3VDC’ cases are very nearly zero for all time, as expected, and the ‘unterminated’ case is also very low.

Next, I added a 0.5V p-p signal at ~520Hz to the sensor input, and re-ran the program.  After capturing the ‘Final Value’ data as before, I added it to the above plot, as shown below

Final Value vs Cycles for 0.5V p-p input

As can be seen in the above plot, the ‘Final Value looks much more reasonable than before. When plotted on the same scale as the ‘grounded’, ‘unterminated’, and ‘+3.3VDC’ conditions, it is clear that the 0.5V p-p case is a real signal.

Then I ran a much longer term (11,820 cycles, or about 22-23 sec) test with 0.5V p-p input, with the following results.

As can be seen from the above plot, the final value is a lot more ‘spiky’ than I expected.  The average value appears to be around 30,000, but the peaks are more like 60,000, an approximately 3:1 ratio.  With this sort of variation, I doubt that a simple thresholding operation for initial IR beam detection would have much chance of success.  Hopefully, these ‘spikes’ are an artifact of one or more remaining bugs in the algorithm, and they will go away once I find & fix them 😉

Update:  Noticed that there was a lot of time jitter on the received IR waveform – wonder if that is the cause of the spikes?

sensor waveform jitter. Note that this display is separately (Vert Mode) triggered.

Following up on this thread, I also looked at the IR LED (transmit) and photodetector (receive) waveforms together, and noted that there is quite a bit of time jitter on the Tx waveform as well, and this is received faithfully by the IR photodetector, as shown in the following short video clip

 

So, based on the above observations, I decided to replace the Trinket transmit waveform generator with another Teensy 3.5 to see if I could improve the stability of the transmit signal.  Since I never order just one of anything, I happened to have another Teensy 3.5 hanging around, and I soon had it up and running in the setup, as shown below

Replaced Trinket transmitter with Teensy 3.5

Transmit and receive waveforms

As the above short video and photos show, the Teensy implementation of the transmit waveform is much more stable than the Trinket version.  Hopefully this will result in better demodulation performance.

The next step was to acquire some real data using a 0.5V p-p input signal through the IR beam path.  I took this in stages, first verifying that the raw samples were an accurate copy of the input signal, and then proceeding on to the group-sum, cycle-sum, and final value stages of the algorithm.

Sample capture using an input of 0.5V p-p through the IR path

GroupSum I/Q plots using 0.5V p-p Input Signal

I then used Excel to compute the cycle sums associated with each group of 4 group sums

Calculated Cycle Sums for a 0.5V p-p Input Signal

And then I used Excel again to calculate the ‘Final Value’ from the previously calculated cycle sum data

Calculated final value

Keep in mind that all the above plots are generated starting with real IR photodector data, and not that large of an input at that (0.5V p-p out of a possible 3.3V p-p).

The next step was the real ‘proof of the pudding’.  I ran the algorithm again, but this time I simply printed out final values – no intermediate stages, and got the following plots

Final Value vs time, for 0.5V p-p Input Signal

Detail of previous plot

From the above plots, I think it is clear that the algorithm is working fine, and most of the previous crappy results were caused by poor transmit timing stability.  I’m not sure what causes the ripple in the above results, but I have a feeling my friend and mentor John Jenkins is about to tell me! 😉

Sleeeeeep, I need sleeeeeeeeeeep….

Frank

 

 

 

IR Modulation Processing Algorithm Development – Part IX

Posted 19 June, 2017

In my last post on this subject, I showed that my 4-sensor band-pass filter (BPF) algorithm was feasible when run on a Teensy 3.5 SBC.  However, what I haven’t done  yet is to verify that the algorithm is indeed producing valid results, when fed with real sensor input.

I should be able to verify proper algorithm operation with my single-sensor test bed (as shown in the following photo) by moving the single sensor input line to each sensor channel (ADC input) in turn, and monitoring the data at different stages in the processing chain.

Teensy 3.5 installed on my algorithm test bed, with the Uno shown for size comparison. The small processor in the foreground is an Adafruit ‘Trinket’

Since I now have plenty of RAM to play with, I should be able to save a representative sample of the input data and intermediate results in suitably sized arrays, run the algorithm long enough to fill those arrays, and then print them all out at the end.

  • I will probably want to run the process long enough to completely fill the 64-element I & Q ‘running sum’ arrays.  These arrays already exist for all 4 sensor channels, so this has no effect on available RAM
  • The next step backward in the chain are the I & Q ‘cycle group sum’ elements (one pair per sensor channel) used to generate one element in the running sum arrays.  To store all these cycle group sum elements will require two 256-element arrays per sensor channel.
  • And the first step in the process is the raw sensor input data.  To store all the data required to generate 64 elements in the running sum arrays will require a single 1280-element array per sensor channel.

In summary, to instrument one sensor channel from start to finish will require

  • 1ea 1280-element array to hold the raw data
  • 2ea 256-element arrays to hold the cycle group sums

for a total of 1280 + 512 = 1796 elements at 2 bytes/element = 3592 bytes.  If I wanted to do this for all 4 sensor channels at once, the total would be 14368 bytes, still well within the 192KB RAM availability on the Teensy – nice!

Results – Capture Stage:

The first step was to capture/display the raw ADC data to make sure that part was operating correctly.  The plots below show all 4 sensor channels.

Raw ADC data for all 4 sensor channels, 1280 elements (enough to fill the entire 64-element running sum array)

First 40 elements of the raw ADC capture

As can be seen in the above plots, channel 1 shows the 520Hz detected IR waveform, and the other three channels show just noise.

Results – Intermediate Stages:

The next step was to verify proper operation of the step that accumulates a 1/4 cycle group of samples and generates the I & Q ‘sample group sum’ components.  To verify this stage of the algorithm, I captured 5 cycles of data, as shown below:

Sensor channel 1 raw data and I/Q sample group sums

In the above plot, the dark blue line is the raw ADC data input, which varies from about the ADC maximum of 4096 to about 3890,  or about 161mV (3.3V ADC reference and IR detector supply).  The resulting ‘sample group sums’ are shown in orange (for the I component) and gray (for the Q component).  The significance of the plot is that the sample group sums and the I/Q component generation appears to be happening correctly.  The orange points follow a {+1, +1, -1, -1} sequence, while the gray ones follow a {+1, -1, -1, +1} sequence, as expected.

Next, I printed out this same 5-cycle segment in text form, as shown below (double-click in code window to enable scrollbar)

The above table shows the raw data, the sample-group sums, and the corresponding cycle-group sums. For example, the first set of 5 data samples adds to 19673.  Since this is the first sample-group sum, it is multiplied by “+1” to form the I component, and “-1” to form the Q component, and these are shown adjacent to the last raw data in that sample group.  After 4 such sample-group sums, the cycle-group sum I/Q components are generated by adding the 4 sample-group I/Q components respectively; for the first cycle these are -1736 & -246 as shown adjacent to the 20th sample (sample #19).

Results – Final Stages:

The cycle-sum I & Q components generated above are saved in separate 64-element circular buffers, and the running sum of these buffers are then used to form the final demodulated value for the channel of interest.  The final value is computed as the sum of the absolute values of the I & Q component running sums, i.e. FV = abs(RunningSumI) + abs(RunningSumQ).  To demonstrate proper algorithm functioning, I printed out the computed final values for well over 1000 cycles of raw data, as shown in the Excel plot below

Computed final values vs complete input data cycles for sensor channel 1

As shown in the plot above, the final value rapidly rises from zero to around 2×106 in the first 64 cycles of the run, after which it generally levels off for the rest of the run.  There is quite a bit of ripple on the signal, which my friend and mentor John Jenkins mentioned might happen as the non phase-locked input and sampling frequencies slowly slid by each other (at least I hope that is what is happening!).

So, it looks like the algorithm is doing what it should, and my ‘scope measurements to date indicate that the Teensy is doing it all without breaking a sweat, even with print statements thrown in.  It appears that I could probably double the number of samples/cycle and still have plenty of time to finish all the computations.

However, there are still a number of things to be accomplished before this new feature makes it into the field.

For starters, I’m not sure how to normalize the final value.  For a fairly weak (~160mv out of 3V) signal the final value is north of 2 million – what happens for stronger signals, and how to I normalize this down to a range that I can use to drive an analog output?  I suppose I could simply apply the IR modulation signal directly to the analog input (bypassing the IR path entirely) and see what happens, but I’d also like to understand the math.  Maybe John Jenkins can help with this (hint, hint, wink, wink!).

Also, I’d like to validate the idea that this algorithm will selectively reject other signals that aren’t close to the desired 520Hz modulation frequency.  I plan to test this by modifying the Trinket algorithm to make it a swept frequency generator (say from 470 – 570 Hz) and see how the output changes.

Stay tuned!

Frank

 

 

 

 

 

IR Modulation Processing Algorithm Development – Part VII

Posted 17 June 2017

In my previous post on this subject, I discussed my decision to change from an Arduino Uno SBC to a Teensy 3.5 for implementing the  ‘degenerate N-path’ digital band-pass filter (BPF) originally introduced to me by my old friend and mentor John Jenkins.  After replacing the Uno with the Teensy and getting everything running  (which took some doing, mostly due to my own ignorance/inability), it was time to see if the change would pay off in actual operation.

In my initial perusal of the available documentation for the Teensy 3.x SBC (have I told you lately how much I love the widespread availability of information on the  inet?), I ran across some new programming features that aren’t available in the rest of the Arduino world.  The Teensy 3.x supports two independent 32-bit timers, supported by two new libraries (TimerOne and TimerThree).  When I first looked at this new functionality, I thought – “wow – this is just what I need to implement the sampling front-end portion of the digital BPF – I can use it with an appropriate ISR to get accurate sample timing!”.   And then I ran across Paul’s ‘Delay and Timing‘ page with it’s description of the new ‘elapsedMillis’ and ‘elapsedMicros’ functions; These functions allow for accurate periodic execution of code blocks inside the normal ‘loop()’ function, without having to deal with interrupts and ISRs – cool!  And then I ran across the ‘FrequencyTimer2’ library written by Jim Studt….

So now I found myself going from no real good options for accurate sample timing to a ‘veritable plethora’ of options, all of which looked pretty awesome – what’s a guy to do?  Since the ‘elapsedMicros’ option looked like the simplest one to implement, I decided to try it first.

elapsedMicros:

From previous work I have a Trinket SBC transmitting an IR beam modulated by a square-wave at approximately 520Hz.  The plan is to sample this waveform 20 times per cycle, and to have the sampling frequency as close as possible to 20×520 = 10.4Ksamples/sec, or approximately 96μS/sample.

I created a small test program to explore the feasibility of using the ‘elapsedMicros()’ function for IR detector sensor sampling.

 

In the above program, I simply generate a 10μS pulse every 95.7μS.  The ‘95.7’ value was empirically determined by watching the transmitted  IR waveform and the 10μS pulses together on a scope, and adjusting the value until the difference between the two frequencies was as small as possible (i.e. when the movement of the transmit waveform compared to the pulse train was as slow as possible), as shown in the short video below:

 

In the above video, the lower trace is the generated pulse train, and the upper trace is the transmitted IR modulation waveform.  The scope trigger was set to the pulse train, with the modulation waveform free to slide left or right based on the ‘beat frequency’ between the two waveforms.

Next, I added code to save ADC samples to an array for later printout.  Now that I am no longer constrained by the minuscule amount of RAM available on the Uno, I opened up the array size to 2000 elements to allow more viewing time before the program was interrupted by the serial output delays.  The code for this and the resulting Excel plot are shown below:

The resulting 2000 element array was dropped into Excel and plotted, as shown below:

All 2000 samples from the test program

First 40 samples. Note that 40 samples covers exactly two cycles of the modulation waveform

So, it looks like the ‘elapsedMicros()’ function is doing exactly what I want it to do – sampling the input waveform at almost exactly 20 samp/sec without me having to figure out the exact delay time needed.

The next step was to determine how much ‘free time’ is left over for other processing steps like sampling multiple sensor channels, doing the ‘sample’ and ‘cycle’ sums, etc.  For this step, I removed the array loading section and replaced it with a call to ‘delayMicros()’.  Then I manually adjusted the delay value until the period of pulse train started expanding away  from the desired 95.7μS value.  The result was that a delay value of 85μS did not change the pulse period, but a value of 90μS did (slightly).  So, I have between 85 and 90μS of ‘free time’ available (out of a total of 96!!!)  for other processing chores.  Adding a single call to ‘analogRead(IRDET_PIN)’ reduced the available ‘free time’ by about 15μS, from between 85 & 90 to between 70 & 75μS.  This shows that the time for a single analog read is about 15μS, which may be due to the same pre-scaling issue as I saw on the Uno (to be determined).  In any case, even if I utilize 4 sensor channels, I should be have about 25μS left over for the summation and array load operations.

To investigate the analogRead() timing issues, I set up a small program to measure the time required to read a pin 1000 times.  Here’s the code:

With the above code, and with all default settings, the time required for 1000 reads was 17mSec, so about 17μS, which tracks well with  the above measurements.

After changing the conversion speed to ADC_CONVERSION_SPEED::HIGH_SPEED, the time required for 1000 measurements was reduced to 11mS, so about 11μS per read.

I ran a whole series of test with the different Teensy ADC library settings, with the following results.  All times are in microseconds, and are the average of 1000 iterations

  • conversion and sampling speed set to “HIGH”: 10.997
  • all adjustments commented out: 17.281
  • just conversion speed set to “HIGH”: 11.014
  • just sampling speed set to “HIGH”: 15.190
  • just resolution changed to 12 bits: 17.276
  • just resolution changed to 8 bits: 17.242
  • HIGH conversion and sampling speeds, and with 8-bit res: 8.931
  • HIGH conversion and sampling speeds, and with 12-bit res: 10.998
  • All of the above, plus averaging set to 1: 4.758

So, I can get the ADC time down to about 5uS/sensor, which means that even with four sensor channels being monitored, I will have over 70uSec for ‘other stuff’, which should be more than enough to get everything done.

Frank

 

IR Modulation Processing Algorithm Development – Part V

Posted  09 June, 2017

In getting the Arduino code working on my Uno/Trinket test setup (shown below), I have been having some trouble getting the delays right.  It finally occurred to me that I should run some basic timing experiments, so here goes:

Sample Group Acquisition Loop:

this is the loop that acquires analog samples from the IR detector, and sums 1/4 cycle’s worth into a single ‘sample group’.  To measure this time, I ran the following code:
int startusec = micros();
int sum = 0;
for (int i = 0; i < 1000; i++)
{
int samp = analogRead(SQWAVE_INPUT_PIN1);
sum += samp;
}
int endusec = micros();
Serial.print("time required for 1000 analog read/sum cycles = "); Serial.println(endusec - startusec);

The time required for 1000 cycles was 15064 uSec, meaning that one pass through the loop takes an average of just over 15 uSec. Adding a 85 uSec delay to the loop should result in a loop time of exactly 100 uSec, and a 1000 pass loop time of 100,000 uSec or 0.1sec.  The actual result was 99504, or about 99.5 uSec/cycle – pretty close!

Next, I replaced the summation with a write to a 500-element array (couldn’t do 1000 and still fit within the Uno’s 2K memory limit), and verified that this did not materially change the loop timing.  The time required for 500 loops was 49788; twice that time would be 99576, or almost exactly the same as the 99504 time for the summation version.

Then I tweaked the delay to achieve as close to 25 complete cycles as possible, as shown in the Excel plot below.  With an 82uSec loop delay, the total time for 500 loop iterations was 48272, or about 96.544 uSec per loop iteration.

96.544 uSec per loop iteration, and 20 loop iterations per cycle gives 20*96.544 = 1930.88 uSec per cycle or 518 Hz.  This is very close to the 525Hz value I got from my O’scope frequency readout when I first fabricated my little test setup.

Next, I coded 500 iterations of a two-detector capture/sum operation, and got: “time required for 2-detector 500 analog read/store cycles = 15520”.  So,  about 31 uSec/iteration, or almost exactly twice the one-detector setup.  A four-detector setup yielded a time of 30352 uSec for 500 iterations, or about 60.15 uSec/iteration.  So, a 4-detector setup is possible, assuming the Uno 2KB memory constraint issue can be addressed successfully.

In summary:

  • It takes about 15 uSec to read each sensor’s A/D value and either sum it or store it in an array
  • A four-sensor setup can probably be accommodated, but only if the required summing arrays fit into available memory (not possible for Uno, but maybe for others.
  • A loop delay value of 82 uSec results in almost exactly 20 samples/cycle.

Stay tuned

Frank

 

 

 

IR Modulation Processing Algorithm Development – Part III

Posted 27 May 2017

In my previous post I demonstrated an algorithm for processing a modulated IR signal to extract an intensity value, but the algorithm takes too long (at least on an Arduino Uno) to allow for 20 samples/cycle (admittedly way over the required Nyquist rate, but…).  So I decided to explore ways of speeding up the algorithm.

First, the baseline:  The starting point is the 17,384 μSec required to process 100 samples in the current algorithm, or 174 μSec/sample.  At an input frequency of 520Hz, 20 samples/cycle is about 96 μSec/sample, so I’m off by a factor of 2 or so.  And this is only for one channel, so I’m really off by a factor of 4 (for a 2-channel setup) or 8 (for my current 4-channel arrangement)

As an experiment, I reduced the running average length from 5 to 1 cycles, or from 100 to 20 samples.  This reduces the shifting operation load by a factor of 5, and resulted in a total processing time of 1876 μSec for all 100 samples – wow!

Then I discovered I had failed to uncomment the line that loads the new running average value into the front of the running average array, so I put that back in and re-ran the measurement.  This time the number came up as 10748 μSec!  This is just not possible!  It is impossible that 10,000 (100 iterations/sample, 100 samples) iterations of a copy operation from one location in the array to another one takes 1/10 the time as 100 iterations (1/sample) of a copy operation from a variable into the array – not possible!!!

But, since it was happening anyway – whether possible or not, I decided I was going to have to figure it out :-(.  So, I changed the line

RunningAvg1[0] = (int)chan1Avg;

to

RunningAvg1[0] = 0;

and re-ran the measurement.  This time the total for processing 100 samples was 1896 μSec – much more believable!  So, what’s the difference between these two operations?  The only thing I could think of is that it must take a lot of time to convert a double to an int.

So, I ran a test where I executed the ‘RunningAvg1[0] = (int)chan1Avg;’ line 10 times, all by itself, and measured the elapsed time.  I got 72 μSec – a much more believable number, but not what I was expecting.  Increasing the number of iterations to 100 resulted in an elapsed time of 672 μSec – consistent with 72 μSec for 10 iterations.  That’s nice, but I’m still not any closer to figuring out what’s going on.

Well, after a bunch more experiments, I think I have the problem narrowed down to the use of floating point math on a few operations.  I have seen some posts to the effect that floating point math is much slower than integer math on Arduino processors, and these experiments tend to bear that out.  I should be OK with integer math everywhere, I hope ;-).

After completely re-writing the algorithm to eliminate floating point math (and correcting several logic errors – oops!), I re-ran the 100-element process for 1 channel, with the following results:

All components – original captured samples, running average, AC component, and full-wave rectified component. Note elapsed time of 3008 uSec

From the above Excel plot, it is clear that the algorithm successfully extracted the full-wave rectified value for the incoming modulated IR signal, and did so in only 3008 uSec for 100 samples.  This should mean that I can easily handle up to three simultaneous channels, and maybe even four – YAY!

Another run with two simultaneous channels was made.  The following Excel plot shows the Channel 2 results, along with the elapsed time for both channels.

Channel 2 all components – original captured samples, running average, AC component, and full-wave rectified component. Note elapsed time of 4268 uSec

The above results for two channels strongly suggests that all four channels in the current hardware implementation can be processed simultaneously while still maintaining a 20 sample/cycle sample rate.  This is extremely good news, as it implies that I can ‘simply’ insert an Arduino Uno or equivalent between the detector array and the robot controller.  The robot contoller will continue to see left/right analog values as before (but inverted – more positive is more signal), but background IR interference will be averaged out by the intermediate processor – cool!

Rather than use a Uno, which is physically very large, I hope to be able to use something like an Adafruit Arduino Pro Micro, as shown below:

Adafruit’s Arduino Pro Micro. 16MHz, 9 Analog 12 Digital I/O

This should fit just about anywhere (probably on top of the sunshade), and be very easy to integrate into the system – we’ll see.

Stay tuned!

Frank

 

 

 

IR Modulation Processing Algorithm Development – Part II

Posted 25 May 2017

One of the things I didn’t understand about the analog sample runs from my previous post was why there were so many cycles of the IR modulation signal in the capture record; I had set the algorithm up to capture only 5 cycles, and there were more than 10 in the record – what gives?

Well, after a bit of on-line sleuthing, I discovered the reason was that the A/D conversion process associated with the analogRead() function takes a LOT longer than a digitalRead() operation.   This put a severe dent in my aspirations for real-time processing of the modulated IR signal, as I would have to do this for at least two, and maybe four independent signal  streams, in real time – oops!

One thing I have discovered for sure in the modern internet era; if you are having a problem with something, it is a certainty (i.e. Prob = 100%) that many others in the universe have had the same problem, and most likely someone has come up with (and posted about) one or more solutions.  So, I googled ‘Arduino Uno faster analogRead()’, and got the following hits:

The very first link above took me to this forum post, and thanks to jmknapp and oracle, I found the Arduino code to reset the ADC clock prescale factor from 128 to 16, thereby decreasing the conversion time by a factor of 8, with no reduction in ADC resolution – neat!

To test the effect of the prescaler adjustment, I measured the time it took for 100 ADC measurements with no delay between measurements.  As shown below, there is a dramatic difference in the ‘before’ and ‘after’ plots:

 

100 ADC cycles with no delay, prescale = 128

100 ADC cycles with no delay, prescale = 16

Next, I adjusted the delay between ADC cycles to collect approximately 5 cycles at the 520Hz input rate, as shown below:

Delay adjusted so that 100 samples ~ 5 cycles at 520Hz.

With the prescaler set to 16, the ADC is much faster.  With a 5-cycle collection window at 520Hz, I have 80 uSec/cycle to play with for other purposes, so it seems reasonable that I can handle multiple input streams with relative ease – YAY!!.

The next step was to simulate a 4-channel capture operation by capturing 400 samples, 100 each from four different channels. In this simulation, all the data comes from the same IR link, but the processing load and timing is the same.  All the samples from the same time slot are taken within a few microseconds of each other, and the loop (inter-sample) delay was adjusted such that approximately five cycles were captured from each ‘channel’, as shown in the following Excel plot

Simulated 4-channel capture

As can be seen in the above plot, the channel plots overlap almost exactly.  What this shows is that the Arduino Uno can capture all four IR detector channels at sufficient time resolution (about 20 samples/cycle) for effective IR signal detection/evaluation, and with sufficient time left over (about 30 uSec) for some additional processing.

If the design is changed from four channels to just two, then the processing load goes down significantly,  as shown in the following plot

Simulated 2-channel capture

To complete the simulation, I added the code to perform the following operations on a sample-by-sample basis:

  • Update the running average of the sample array
  • Subtract the running average from the sample, and take the absolute value of the remainder (full-wave rectification)
  • Store the result in another array so it can be plotted. This last step isn’t necessary except for debugging/evaluation purposes

Initial results as shown below are very promising. The following Excel plots show the results of processing 100 ADC samples in real time.  First 100 samples were loaded into an array to represent the last 100 samples in a real-time scenario, and the running average value was initialized to the average of all these samples.  Then each subsequent real-time sample was processed using the above algorithm and the results were placed in holding arrays for later printout, with the following results

All components – original captured samples, running average, AC component, and full-wave rectified component

Detail view of original captured samples and the running average component

Detail view of the AC component of the original captured samples and the computed full-wave rectifed component

The above plots confirm that the ADC samples can indeed be processed to yield the full-wave rectified intensity of a modulated IR beam.  However, there is a fly in the ointment – it takes too long; it took 17,384 μSec to process 100 samples – but 100 samples at 20 samples/cycle only takes approximately 9600 μSec – and this is only for one channel :-(.  I will need to find some serious speedup tricks, or reduce the number of samples/cycle, or both in order to fit the processing steps into the time available.

Stay tuned,

Frank

 

 

 

 

 

 

IR Modulation Processing Algorithm Development – Part I

Posted 23 May 2017

As you may recall from previous posts, I have been collaborating with my long-time friend and mentor John Jenkins on the idea to use square-wave modulation of the charging station IR beam to suppress ‘flooding’ from ambient IR sources such as overhead incandescent lighting and/or sunlight.

More than likely, If it is possible to implement a software processing algorithms to recover steering information from potentially corrupted data, it will have to be housed on a dedicated processor.  So, I decided to set up a separate test setup using two processors – one to generate a square-wave modulation waveform, and another to receive that waveform through an IR link.  The link can then be modified in a controlled way to simulate link losses and/or ‘flooding’.  The initial hardware setup is shown below.

Initial test bed verification using 12cm separation

Scope shot showing transmitted and received waveforms, 12cm separation

Algorithm test bed with IR link range set to 78cm

Then I ran my little test program on the receiver processor that simply acquires 100 samples at roughly 20 samples/cycle and then prints out the results.  The following two images are Excel plots of the results for 12cm & 78cm separation.

As can be seen from these two plots, the 12 & 78cm separation values provide a reasonably good simulation of the ‘very good’ and ‘reasonably crappy’ signal conditions.

Next I verified that I can successfully ‘flood’ the receiver with my portable battery-operated IR signal generator.  I monitored the transmitted and received waveforms, without and then with flooding.  In both cases, the bottom trace is the 5V square-wave transmitted signal, shown at 2V/div, and the top trace is the received signal shown at 1V/div.  The ground for both traces is the same line on the scope screen.

78cm separation, no flooding signal. Bottom trace is transmit @ 2V/cm, top is receive @ 1V/cm, ground for both is same line

78cm separation, with flooding signal. Bottom trace is transmit @ 2V/cm, top is receive @ 1V/cm, ground for both is same line

Applying flooding signal with battery-operated IR signal generator

As can be seen in the scope photos, I can indeed produce almost 2V of ‘flooding’ using the IR signal generator, so I should be able to determine whether or not a particular recovery algorithm is successful at suppressing flooding effects.

Stay tuned

Frank

 

 

 

 

Charging Station System Integration – New Sunshade Testing

Posted 19 May 2017

While working with John Jenkins on the modulated IR beam idea, I decided to run some tests with the current 4-detector design, to see how the new sunshade with center divider was affected by ambient IR on a bright sunny day.  So, I disabled Wall-E2’s motors and then placed it at several different critical spots in the entry hallway.  At each location I used my little IR test generator to mark the beginning and the end of the test for that location, and then moved on to the  next one.  The locations are shown in the following photos, in the order that the tests were run.

Position 1: Near where Wall-E2 transitions from wall-tracking to IR beam homing

Position 2: This is where Wall-E2 has been winding up when it homes on the outside sunlight instead of the IR beam

Position 3: Here Wall-E2 should be firmly fixated on the IR beam

These results, combined with my earlier IR response tests with a single phototransistor are encouraging, because it is clear that at least in this case, Wall-E2 should have no difficulty discriminating between the ambient IR and the charging station IR beam.

I ran some homing tests, which Wall-E2 handled with ease; unfortunately God had already turned the lights out on this side of the world, so the tests weren’t in the presence of daylight IR interference.  I’ll do some more real-world discrimination testing tomorrow, and I am hopeful that the new sunshade-with-divider version will be successful, at least for this part of the house.

Stay tuned,

Frank

 

 

IR Light Follower for Wall-E2, Part XI – Center Divider Investigation

Posted 16 May 2017

One of the suggestions John Jenkins made during his visit, in addition to the idea of modulating the IR beam to suppress ‘flooding’ from ambient IR sources, was the idea of placing an opaque divider between the left and right halves of the detector array.  He thought that I might even be able to reduced the detector array from four to two phototransistors and still get good homing performance, assuming that each of the two detectors had sufficiently wide beamwidths to accommodate off-axis IR beam intercepts.  The current detectors have a +/- 12º beamwidth, but are arrayed in such a way as to provide well over 60º aggregate coverage.  To do the same thing with just two detectors would require parts with considerably wider beamwidths.  the TAOS TSL267 (another one of JJ’s suggestions) has an approximately +/-30º beamwidth at the half-response points, so they seem almost ideally suited for this application.  As a major added bonus, the TAOS parts feature a photo-diode integrated with an op-amp to address the dynamic range issue mentioned by John.  The diode operates in its linear range, and the op-amp amplifies the IR signal to useful levels.  The only fly in the ointment is that the op-amp gain isn’t adjustable, and its output limits at fairly low light levels – bummer!

In order to investigate the opaque divider idea, I decided to run some bench angular response tests using my robot’s 4-detector array with and without a center divider.  I printed out a copy of the compass rose graphic I had hanging around from my magnetometer project (one of my more spectacular failures) and set up a bench test with Wall-E2 and my little IR test source, as shown below.

Angular response test setup

Closeup of the ‘sunshade’ cowling (black rectangular opening) around the 4-detector array

Closeup of sunshade with divider installed

The results without the center divider were about as expected, with about +/- 45º coverage, as shown in the Excel plot below

Response vs angle for 4-detector array, without center divider

+/- 60 deg response vs angle for 4-detector array, without center divider

With the divider, the response appears to be about the same (note here that the ‘with divider’ response is flipped left/right from the ‘no divider’ case – oops!)

+/- 30 deg response vs angle for 4-detector array, without center divider

Note in the above ‘detail’ view that the response curves for the two center detectors (DET2 & DET3) seem to be very symmetric about the 0º point, as expected.  What this also shows is that just these two detectors could probably be used for homing, if off-axis beam detection weren’t a consideration.

+/- 30 deg response vs angle for 4-detector array, with center divider

The ‘with divider’ plot above shows a significant difference from the ‘no divider’ plot, right at the center.  In the ‘no divider case, the DET1 & DET2 responses are very nearly the same, but in the ‘with divider’ case they are significantly different, and almost perfectly anti-symmetric about the center.  This should allow more precise detection of small left/right deviations from the beam centerline, and therefore more precise homing.

Stay tuned!

Frank

 

 

IR Phototransistor Sensitivity/Dynamic Range Study

Posted 05/14/17

In a previous post, I mentioned that John Jenkins, mentor and old friend, had some ideas regarding my Wall-E2 robot’s problems with homing in on an  IR beam in the presence of ambient IR sources like overhead incandescent lighting and/or sunlight streaming in through windows and doors.   John pointed out that this was really a system dynamic range issue, and it was likely that, as currently configured, the IR phototransistors were running out of dynamic range (saturating) well before the 10-bit A/D’s on the AtMega SBC.  In order to get the detector sensitivity up to the  point where Wall-E2 could ‘see’ the IR beam far enough (1.5-2m) away to avoid hanging up on the lead-in rails (see this post and this post for details), I had to use a very high value collector resistor (330K), which reduces the dynamic range significantly.

In a subsequent email conversation, John suggested that the proper way to handle this problem was to reduce the collector resistor to the point where the detector doesn’t saturate under the worst case ambient IR conditions, and then add amplification as necessary after the detector stage to get the required sensitivity.  John’s point was that as long as the detector response is relatively linear (i.e., it’s not saturated), then there shouldn’t be any loss of information through the stage, so a later linear amplification stage will allow the desired IR signal to be detected/processed even in the presence of interference.  However, if the detector stage saturates due to interference, then it’s basically ‘game-over’ in terms of the ability to later pull the desired signal out of the noise.

This wasn’t exactly what I wanted to hear, as adding the required post-detector amplification stage wasn’t going to be particularly easy – there’s not much left in the way of free real-estate on Wall-E2’s main platform (there’s plenty of room on the second level, but putting stuff there requires inter-level cabling and is a major PITA).  As I was thinking this, I had a flashback to similar conversations with John from 40-45 years ago, when we were both design engineers in a USG design lab; the usual outcome of such ‘conversations’ (to the uninitiated these might be mistaken for shouting matches) was that my circuit got ‘simplified’ by the addition of 50% more parts – but could then withstand a nuclear attack in the middle of a snowstorm at the South Pole!

Anyway, back in the present, John suggested I run some experiments designed to determine the ratio of IR field intensity to collector current for a the phototransistors I was using, so the proper collector resistor value to be computed to utilize the available detector dynamic range without driving it into saturation.  His suggestion was to not  use a collector resistor at all, but to simply connect the collector to +VDC through a current meter, and then expose the detector to worst-case ambient conditions.  I didn’t particularly like this idea, because I only have a manual non-recording multimeter, and I wanted to record the data for later analysis.  So, I decided to program up one of my small SBC’s with analog input capability (in this case, a Pololu Wixel), and set up to measure the voltage drop across a 10 Ω precision resistor in the collector circuit.  The hardware setup is shown in the following photo

Hardware setup for IR Phototransistor Response Experiment

Before and after collecting ‘field’ data, I made a collection run on the bench, using my IR LED test box, with the following results.  As can be seen in the plot, the maximum voltage drop across the 10Ω resistor was approximately 200mV, or about 20mA.

IR detector response bench test. Nose-to-nose with my IR test source

Bench test with IR LED test generator

 

To collect the ‘field’ data, I placed the SBC in different locations around the house, and recorded A/D values.  For these experiments I was using an example Wixel app that printed out the A/D values for all 6 analog inputs, scaled such that the maximum reading was equivalent to the SBC board voltage (3.3V) in millivolts.  In other words, the maximum A/D reading was approximately 3375, or about 3 mV/bit

Location 1: Looking out glass doors on south side of house

Location 2: Direct sunlight through window on south side of house

Location 3: Kitchen floor, looking toward entrance hall and atrium

Location 4: Entry hallway, looking toward atrium. This is the usual starting location for charging station homing tests

Location 5: Entry hallway near door to atrium

These results were not at all what I expected.  Either there is something wrong with the experimental setup, or the ambient light IR field intensity isn’t anywhere near as strong as expected.  If the data can be believed, the ambient conditions are significantly weaker than the bench-test conditions, which were basically nose-to-nose with my IR test LED.  So, if my planned double-check of the hardware doesn’t find any problems, then I’ll change the resistor value from 10Ω to 100Ω and repeat the tests.

17 May 2017 Update

After thoroughly reviewing the hardware setup, I concluded that everything was working properly, but that the 10Ω load resistor simply didn’t provide sufficient drop for reliable measurements.  So, I changed out the 10Ω resistor for 100Ω, and repeated the tests, with the following changes:

  • As before, I started the run by performing a ‘nose-to-nose’ test to verify proper operation, but this time I measured the detector current using a multimeter. The  result was about 15mA maximum current.
  • At each of the 5 locations, I started with a short nose-to-nose section (not shown in the plots) to make sure the detector was operating properly
  • At location 2 (the direct sunlight location), I physically oriented the detector for maximum response.
  • After location 5, I placed the detector on the charging station’s IR beam reflector boresight at about 0.5m distance, and physically oriented the detector for max response.
  • I calculated the detector current for each reading, and plotted that rather than the raw reading.  To calculate the detector current for each measurement, I subtracted the detector reading from the corresponding 3.3VDC supply voltage measurement and divided by 100.

The results of these tests are shown in the plots below:

 

These results are pretty interesting.  As I’m sure John would point out, the direct sunlight response of about 0.2 mA is about 20 times the value required to saturate the phototransistor with the current 330KΩ.  No wonder I was having interference problems – oops!

Stay tuned!

Frank